Form: Steam Quality Test Record (EN 285 Triad)

A plug-and-play record for the EN 285 steam-quality triad: non-condensable gases, dryness value, and superheat at the point of use, with field definitions, acceptance columns, retention, and a filled specimen row.

Document type: Form

Read and copy the template below into your own quality system. It is a generic starting point for your own internal use, provided as is, with no warranty; see the Terms and License. Adopting it does not by itself create compliance.

This is a ready-to-use record for a steam-quality test using the EN 285 triad: non-condensable gases, dryness value, and superheat, measured with a steam-quality test rig at the point of use. Condensate chemistry tells you the steam is clean; this test tells you it can sterilize, and the two are separate. Replace <<FILL: ...>> placeholders, read the acceptance limits off the current edition of EN 285, and file the completed record with the qualification or routine monitoring package. A filled specimen row follows. Confirm the standard against the current source before you rely on it.

Control header

FieldEntry
Document titleSteam Quality Test Record (EN 285 Triad)
Document number<<FILL: e.g. FRM-STM-002>>
System / generator ID<<FILL>>
Point(s) of use tested<<FILL>>
Test rig ID (calibrated)<<FILL>>
Reason for test<<FILL: qualification / periodic / post-maintenance / investigation>>
Retention<<FILL: e.g. life of the system plus records-retention period>>

Field definitions

FieldFormatRequiredWho entersWhen
Date / timeDate-timeYesTesterAt test
Point of useText / IDYesTesterAt test
Non-condensable gases (NCG) result% v/vYesTesterAt test
NCG acceptance limit% v/v (from current EN 285)YesTesterAt test
NCG verdictPass / FailYesTesterAt test
Dryness value resultFractionYesTesterAt test
Dryness acceptance limitFrom current EN 285 (note tighter value for dense metal loads)YesTesterAt test
Dryness verdictPass / FailYesTesterAt test
Superheat resultdeg C rise on expansion to atmosphericYesTesterAt test
Superheat acceptance limitdeg C (from current EN 285)YesTesterAt test
Superheat verdictPass / FailYesTesterAt test
Supply pressurebar / kPaYesTesterAt test
Overall verdictPass / FailYesTesterAt test
TesterName / signatureYesTesterAt test
Reviewer (QA/SME)Name / signature / dateYesReviewerOn review

Instructions

  1. Use a calibrated steam-quality test rig; record its ID and confirm calibration is current before testing.
  2. Test at the point of use, not the generator outlet; steam quality degrades along the distribution.
  3. Read the acceptance limits for NCG, dryness, and superheat off the current edition of EN 285. The dryness minimum is tighter for dense metal loads; apply the value that matches your load.
  4. Record each of the three results, its limit, and its verdict. A single Fail among the three fails the overall test; investigate before use.
  5. Have QA or the SME review and sign. File with the qualification package (for a PQ point) or the routine monitoring record.

Test record

Date / timePoint of useNCG resultNCG limitNCG verdictDryness resultDryness limitDryness verdictSuperheat resultSuperheat limitSuperheat verdictOverallTester
<<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>><<FILL>>

Acceptance criteria

The steam quality passes at a point of use when all three attributes are within their current EN 285 limits at that point. Any single Fail fails the test and triggers an investigation (feedwater, generator separation/carryover, trap performance, or distribution) before the steam is used for sterilization.

References

EN 285 (steam sterilization, large sterilizers) for the non-condensable gas, dryness, and superheat limits and the test method. EU GMP Annex 1 (2022) for the expectation that steam used for sterilization is of suitable quality. Related reading: clean utilities qualification, sterilization validation: moist heat.

Confirm the current version and clause numbers of each reference before issue.

Revision history

VersionDateAuthorSummary of change
<<FILL: 1.0>><<FILL>><<FILL>>Initial issue.

Filled specimen

The following shows one completed row for an example pure-steam point of use serving an autoclave, so you can see how the three verdicts combine. The values are illustrative; take real limits from the current EN 285.

FieldEntry
Date / time22 July 2026, 09:40
Point of useAutoclave AUT-02 chamber inlet
NCG result / limit / verdictcomfortably under the EN 285 cap / per current EN 285 / Pass
Dryness result / limit / verdictabove the minimum / per current EN 285 (standard load) / Pass
Superheat result / limit / verdictwell under the cap / per current EN 285 / Pass
Supply pressure<<FILL>>
Overall verdictPass
Tester / ReviewerJ. Okafor / R. Gomez (QA), 22 July 2026

In this example all three attributes passed at the point of use, so the steam is fit to sterilize. Had the NCG result exceeded the cap, the overall verdict would fail even with dryness and superheat in range, because air pockets shield the load and defeat sterilization. Recording all three, at the point of use, with the load-appropriate dryness limit, is what makes the record defensible.

Common inspection findings this form prevents

  • Steam judged acceptable on condensate chemistry alone, with no EN 285 triad record.
  • Testing at the generator outlet instead of the point of use.
  • Using the standard-load dryness limit for a dense metal load that needs the tighter value.
  • A single failed attribute buried under an overall “acceptable,” with no investigation.
  • An uncalibrated or unrecorded test rig, so the numbers cannot be trusted.

How to adapt this form

  1. Set your document number, generator ID, and test-rig ID in the header.
  2. Enter the EN 285 acceptance limits from the current edition, matched to your load type.
  3. Record every test at the point of use with all three verdicts.
  4. Route any Fail through your deviation process before the steam is used.
  5. Confirm the standard against the current published version before issue.