This is a ready-to-use addendum checklist. It does not replace the general batch disposition package review; it adds the device-constituent items a drug-only checklist does not ask for. Complete it alongside checklist: batch release disposition package review for any drug-led or biologic-led combination product. Mark each item Pass, Fail, or NA, record the evidence reference, and route any Fail per your deviation procedure before the disposition authority signs. Replace every <<FILL: ...>> placeholder. A filled specimen follows.
How to use this addendum
- Complete one addendum per batch, tied to the same disposition record as the general package checklist.
- A Fail on any item here blocks release exactly as a Fail on the drug side does; the device constituent is not a lesser check.
- The reviewer confirming device items should be able to read a functional test result, not only a drug cGMP record; name the qualified reviewer in the header.
- If the product is device-led rather than drug-led, confirm with Regulatory Affairs which base quality system and disposition framework actually applies before using this addendum.
Section D: Device constituent
Disposition conclusion for this addendum
Filled specimen
Header. Product: prefilled autoinjector, biologic-led combination product. Batch: AIX-2026-0091. Device-constituent reviewer: M. Oyelaran, device quality engineer. Review date: 2026-09-30. Linked disposition record: DISP-2026-1144.
Conclusion: all Section D items Pass or NA. Device-constituent reviewer sign-off: M. Oyelaran, signed, 2026-09-30. Disposition proceeds using both this addendum and the general disposition package checklist.
Common findings this addendum prevents
- A batch released on drug-side results alone, with no device functional test result ever entering the disposition package.
- Device component lot traceability maintained separately from the batch record and never actually checked at disposition.
- A device-attributable complaint trend recognized by Quality but never cross-checked against the batch being dispositioned.
- The assembled product’s configuration drifting from the design output referenced in the marketing application, with no one dispositioning against the current design output record.
How to adapt this addendum
- Tailor Section D’s functional-test item (D1) to the actual device type: an inhaler checks dose uniformity and actuation count differently than an autoinjector checks actuation force and injection time.
- If the combination product is co-packaged rather than single-entity, add an item confirming the packaging-step reconciliation between the two constituent manufacturing histories.
- Route this addendum to the same disposition record ID as the general package checklist so an inspector sees one linked file, not two disconnected reviews.